8ch Coil Configure Part Per Upgrade

2026-02-27 - Relinque mihi nuntium

SYSTEM: 1.5T Excite II (Software Version. 11.0M4)

PROBLEMA / SYMPTOM

1. Systema 10.0M4, post upgrade ut Excite II 11.0M4, 8 canalis NV et corpus scan non possunt

2.Error nuntius: Coil praescriptum RF coil non congruit lectus.

Troubleshooting HISTORIA / INDICIUM

1. 8ch NV et corpus coil remove a 'site coil', programmatum sileo, adde 8ch NV et corpus e 'coil' sito coil', programmatum sileo, 8ch NV est OK, sed idem problema occurrit cum corpore 8ch.

2. Test 8ch corpus in alio situ (Excite II), scan est OK.

3.LFC pro Shanghai PLA 455, adhuc non operatur

4.Check error log: Coil praescriptum RF non aequare. Coil connect status: 0, coil coilis superior: Nulli, coil inferioris status connect: 1, coil inferior coil: 7t8jkd98cg58919otdqaft7c0bwilvc8, Rx coil nomen: 8Ch Corpus inferioris, Rx coil code: 7t8jkd98cg58919otdqaft7c0bwi1vc8

5. Edit 8ch superior, inferior et plena FOV coil' 7t8jkd98cg58919otdqaft7c0bwi1vc8' ut '7t8jkd98cg58919otdqaft7c0bwilvc8'. Sileo software, scan est OK!

RADICES CAUSAE / SOLUTIO

Quia 8ch NV et coil corpus in 10.0M4. Haec duo volumina configuratio manually in Excite I editum est, figura "coil codicem" non habet. Post upgrade, coil figurae restitutae idem cum 11.0 M4.

Mitte Inquisitionem

X